Plastometer PM-04
Measurement of plasticity of coal is performed according to procedures described in norm ASTM D 2639 - 90. During the test, a dependence of rotation of agitator, which is placed in a retort with a measured coal, on temperature is measured. During a measurement, the agitator is exposed to a constant torque, which is exerted in a propeller unit by a hysteresis brake.
Tests are performed in a temperature range from 300 to 600 ˚C. To comply with the measurement procedure, it is necessary to follow all the steps defined in the norm ASTM D 2639 - 90.
The measurement is controlled automatically using an enclosed software.
Furnace
Maximal temperature | 600 ˚C | |
Thermocouple | type K | |
Dimensions | inner diameter | 66 mm |
depth | 170 mm |
Measuring head
Maximal torque | 300 g cm | |
Hysteresis clutch insensitivity | < 2 g cm | |
Rotation sensor | optoelectronical | |
Rotation sensor transfer | 100 imp/rev |
Device module
Input voltage | 230 V | |
Wattage | 800 VA | |
Temperature controller | e3508 | |
Regulatory thermocouple | type K | |
Measurement duration | 2 hod | |
Comunication interface | ETHERNET | |
Control | automatic | from SW |
Weight | 35 kg |
This product was probably delivered to Czech Republic only so there is no documentation available in English for now. If you have any questions about this product, we are ready to answer any of your questions.